This paper describes a feasibility study of a novel
force sensor that can be used to localise tissue abnormalities
and provide tactile feedback to the surgeon during minimally
invasive surgery. The proposed sensor makes use of an airsupported rigid ball mounted at the end of a tubular shaft to
indent the tissue under investigation. Variations in tissue
stiffness which cause changes in position of the ball are
measured by using an optical sensing scheme. The sensor
enables rapid acquisition of tactile information over large areas
of soft tissue. Laboratory experiments were conducted to
demonstrate the feasibility of the proposed sensor system. The
outcome of the conducted experiments shows similarity to the
results from a cylindrical wheel-based force sensor [1] which
are shown here for comparative purposes.
Description
Keywords
Citation
D. Zbyszewski et al., "Wheel/tissue force interaction: A new concept for soft tissue diagnosis during MIS," 2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Vancouver, BC, Canada, 2008, pp. 5556-5559