Name: | Description: | Size: | Format: | |
---|---|---|---|---|
986.27 KB | Adobe PDF |
Advisor(s)
Abstract(s)
The design and operation of two independent vibrating-wire viscometers are described. The instruments are intended for operation in the liquid phase at pressures up to 300 MPa and have been designed specifically for this purpose using the detailed theory of the device. Extensive evidence is adduced to
demonstrate that the operation of the viscometers is consistent with the theory.
Although the instruments attain a precision in viscosity measurements of __+1% when used in an absolute mode the accuracy that can be achieved is no better than __+3%. However, if the instrument is calibrated for two welldefined instrumental parameters, the uncertainty in the reported viscosity is
improved to _+0.5%. The results of measurements of the viscosity of normal
heptane in the temperature range 303 to 348 K at pressures up to 250 MPa made with one of the viscometers are reported. The results are shown to be totally consistent with measurements reported earlier using the instrument designed for lower pressures.
Description
Keywords
High pressure Vibrating wire Viscosity
Citation
Oliveira, Carla Padrel de [et al.] - Vibrating-wire viscometers for liquids at high pressures. "International Journal of Thermophysics" [Em linha]. ISSN 0195-928X (Print) 1572-9567 (Online). Vol. 13, nº 4 (1992), p. 593-615